Metrohm 877 Titrino plus User Manual
Page 33

■■■■■■■■■■■■■■■■■■■■■■
5 Operation
877 Titrino plus
■■■■■■■■
23
"#" stands for a sequential number that you must enter manually. Exam-
ple: if you apply the variable EP# in the formula, only EP is entered. You
will still need to enter the number yourself.
The meanings of the editing functions are explained in chapter 5.2.4,
page 21.
Variables
Pressing Var displays a list with additional variables. You can enter these
variables either directly into the formula or also by selecting them from the
list and applying them with [OK].
Variable
Description
MIM
Initial measured value, i.e. measured value prior to the
processing of the start conditions
MSM
Start measured value, i.e. measured value after the pro-
cessing of the start conditions
MCV
End volume, i.e. total dosed volume at the end of the
titration
ET#
Temperature at endpoint EP# (# = 1…9)
EM#
Measured value of endpoint EP# (# = 1…9)
ED#
Time at endpoint EP# (# = 1…9)
MSV
Start volume
MEN
Electrode zero point pH(0)
MSL
Electrode slope
DD
Duration of the entire determination
MST
Start temperature
MCT
End temperature
FT#
Temperature at fixed endpoint FP# (# = 1…9)
FM#
Measured value of fixed endpoint FP# (# = 1…9)
FD#
Time at fixed endpoint FP# (# = 1…9)
For Molw, see the following section.
Calculation templates
Pressing Templates displays a list with calculation templates. You can
apply these templates directly with [OK].