Figure 12, Installing accessories for the titration – Metrohm 814 USB Sample Processor User Manual

Page 29

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

814 USB Sample Processor

■■■■■■■■

21

6.0xxx.xxx

6.1236.020

6.2740.020

6.2709.070

6.1543.200
6.1543.170

1

1

1

2

2

4

3

5

6

Figure 12

Installing accessories for the titration

1 Insert three spray nozzles (6.2740.020) or rinsing nozzles

(6.2740.030) into the oblique bore holes, see figure.

2 Insert two 6.2709.070 guide sleeves into the titration head, see

figure.

3 Insert a 6.1543.200 titration tip (with antidiffusion valve) into a

guide sleeve.

4 Insert a 6.1543.170 aspiration tip into a guide sleeve.

5 Slide a 6.1236.020 SGJ sleeve with standard ground joint onto the

electrode to be used. Insert the electrode into the titration head, see
figure.