Safety pre cau tions i, Pur pose and func tion 1, Di men sions and weight 1 – Bird Technologies 6730B Series User Manual
Page 9: Power and util ity re quirements 1, Purpose and function, Dimensions and weight, Power and utility requirements, Introduction

Chapter 1
Introduction
This instruction book is intended for use by operators of
the Models 6732B, 6734B, 6734B030, 6735-300A, 6736A,
6736-030A, 6737A, and 6737-030A Termaline Wattmeter.
This chapter contains introductory information including
product specifications, items supplied, and accessory items
available.
Purpose
and
Function
A Series 6730B Termaline Wattmeter is a portable direct
reading absorption wattmeter that both measures and
dissipates RF power. It is a self-contained unit that re-
quires no outside power source or additional equipment,
except for the Models 6737A and 6737-030A which must
have auxiliary water cooling. The load portion forms an
accurate, dependable and practically non-reflective 50
ohm termination for the adjustment, standby and testing
of transmitters under nonradiating conditions.
These Termaline Wattmeters can safely measure and ab-
sorb their maximum rated RF power. The low frequency
units will measure power to within ±5 percent of full
scale over the frequency band of 2 to 32 MHz. The rest of
their frequency range permits power measurement to ±10
percent of full scale. The high frequency units measure to
±5 percent of full scale from 25 to 512 MHz and ±10 per-
cent of full scale from 512 to 1000 MHz. Each meter has
three switch selectable scales and is housed separately
from the load.
Dimensions
and Weight
The dimensions of the individual loads are given in the
specifications on page 3. The meter housings are 5 9/16
inch W x 6 ½ inch H x 4 11/64 inch D (141 x 165 x 106
mm) for the Model 6735, 6736, and 6737 series and 3 7/8
inch W x 5 5/8 inch H x 3 7/8 inch D (98 x 143 x 98
mm) for the Model 6732 and 6734 series. The total
weight for each wattmeter is also given in the
specifications. The shipping weights will run approxi-
mately 20 percent higher.
Power and
Utility
Require-
ments
The Series 6730B Termaline Wattmeters require no exter-
nal source of electrical power other than the RF input
power for operation. The Models 6737A and 6737-030A
do, however, depend on a supplemental water supply for
cooling.
1