Run Up/Coast Down Extension Module Chapter 2

Spacing

Controls the spacing between collected spectra:

Δ

rpm is a fixed speed change between spectra.

%

Δ

rpm is a percentage speed change between spectra.

Δ

Time is a fixed time interval between spectra.

Δ

rpm +

Δ

Time is a combination of speed and time interval settings.

If speed stops changing then spectra are calculated using the Time
Interval value.

%

Δ

rpm +

Δ

Time is a combination of speed and time interval

settings. If speed stops changing then spectra are calculated using the
Time Interval value.

Δ

rpm

%

Δ

rpm

Δ

Time

Δ

rpm +

Δ

Time

%

Δ

rpm +

Δ

Time

Your selection affects the Speed Interval
and Time Interval settings.

Speed Interval (RPM)

Sets the speed interval used in calculating FFT spectra.

This parameter not used when Spacing is set to

Δ

Time.

1…240 rpm or % rpm

Time Interval (sec)

Sets the time interval used in calculating FFT spectra.

This parameter is not used when Spacing is set to

Δ

RPM and %

Δ

RPM.

.001…604800 s

Display Units

Sets the Y-axis units.

The available units depends on the transducer used to collect data.

See

Run Up/Coast Down Setup Parameters on page 18

.

LF Cutoff

Determines whether to apply a low frequency filter (0.3 times the running
speed) to the data.

Off

On

Detection

Sets the signal detection and scaling display for the table:

Peak - Measured RMS value multiplied by the square root of two

(1.4142).

Pk-Pk - Measured RMS value multiplied by two times the square root of

two (2.8284).

RMS - Root mean squared signal level of the input signal.

RMS

Peak

PkPk

Num Traces

Sets the number of traces or order tracks to be displayed in the table.

Your selection determines the available number of Trace Order parameters.

1

2

Trace Order #1-#4

The order numbers to be tracked and displayed in the table.

This value cannot exceed the Max No Orders value set in the RUCD Setup
screen.

See

Run Up/Coast Down Setup Parameters on page 18

.

.001…166 orders

Table 6 - Table Parameter Descriptions

Parameter Name

Description

Values/Comments