Installation/assembly, Lift for re510 – Yamato Scientific RE510 Evaporators, Rotary User Manual
Page 20
![background image](/manuals/576642/20/background.png)
18
Installation/Assembly
Lift for RE510
18. Manual lift (Supplemental function of RE510)
Adjust the lift by using the two knobs located on the right side of the main assembly.
Lift Knob—lift position knob (larger knob)
Lower Limit Knob—sets a lower limit to prevent glass breakage (smaller knob)
Never operate the larger knob without motor or glass assemly.
1. Loosen the smaller knob (lower limit knob). Picture 1
2. Turn and hold the larger knob to “release”, you may now move the position of the lift.
3. Once you have determined the position of the lift return the larger knob to “Lock”. Picture 2, 3
4. After you determine the lowest position, fasten the small knob. Picture 4, 5
The lift will not go lower than the fixed position. However, this function effectively works only when
the small knob is positioned within 5.3 inches ( 135 mm ) from the bottom.
Picture 1
Picture 2
Picture 3
Picture 4
Picture 5