background image

10

Installation/Assembly

RE210

4. Attach the motor to the motor support clamp.

1. Fit the motor bearing bar to the motor support clamp, and fasten the clamping screw tightly.

Put D-cut surface ( flat surface ) of the bar perpendicularly to the screw.

2. Slant the motor to the right (about 45

°) and finger tightly the motor screw.

If you do not fasten the screws tightly,

vibration may occur preventing accurate

measurements or the motor may fall causing

the glass apparatus to break.

5. Attach the controller to the support rod.

Attach the controller to the support rod by attached clamping screw.

Bearing bar

Motor Support Clamp

Clamping Screw

Controller

Motor Screw