Service procedures, Bulb replacement – UVP BioLite User Manual
Page 11

Manual BioLite MultiSpectral Light Source
11
Service Procedures
Bulb Replacement
The BioLite’s Type EKE 150 watt halogen bulb has a standard life of 200 hours. It must be replaced with the
same bulb type only. The UVP part number is 34-0088-01.
The BioLite bulb replacement procedure is as follows:
1. After allowing the unit sufficient time to cool after last use (at least 1/2 hour), unplug the BioLite
from its power source.
2. From the bottom of the unit, remove the brass thumb nut and remove the bulb access door.
3. Remove the bulb using the mechanism built into the bulb holder. Grasp the chrome wire to the
left of the bulb and push it in the direction of the base of the bulb. The bulb will pop out of the
connector.
4. Return the bulb removal mechanism to its original, downward position.
Brass Thumb Nut
Bulb Access Door