beautypg.com

RKI Instruments GX-2012 User Manual

Page 51

background image

Viewing, Printing, Exporting, & Deleting Data in the Data Window • 47

To view and perform desired operations with the interval trend files:

1. With the program already launched, click the Data control button along

the right side of the program window. The Data Window will appear.

2. If necessary, double-click the GX-2012 or Gas Tracer icon in the top of

the Data Window’s upper left frame to see the folders of downloaded
data.

3. Find your instrument by serial number, then click the expanded view

symbol (+) of or double-click the serial number folder to view the
contents.The top folder is untitled and contains the Calibration History,
Bump Test Data, Alarm Event, and Trouble Event folders. The rest of
the folders contain folders for the interval trend files, bar hole mode
data, snap log data, and alarm trend files and are named and
organized by date (month/year).

4. Click the expanded view symbol (+) of or double-click the dated folder

whose contents you want to see.

5. Click on the Interval Trend Folder. In the upper right frame of the Data

Window, a list of file names will appear in the Name column. A prefix of
“iv” indicates an interval trend data file.

The serial number, station ID, user ID, and date/time appear on the line

Figure 43: Data Window - Selecting Interval Trend Data Files

click to

show

Click to

expand

folder

Click to

Interval

trend

data

folders

show

interval

trend

files

files

Double