beautypg.com

RKI Instruments 12 VDC Beacon 200 User Manual

Page 21

background image

12 VDC Beacon 200 Operator’s Manual • 17

5. Connect the wires from the recording device to the recorder output

terminals as shown in Figure 9.

Connecting RKI Detector Heads

Perform the following procedure to connect an RKI detector head to the 12
VDC Beacon 200.

1. Turn off or unplug power to the 12 VDC Beacon 200.

2. Open the 12 VDC Beacon 200 housing door and place the power

switch in the off position.

3. See the detector head instruction manual for instructions on how to

connect wires to the detector head.

4. Install an appropriately rated cable bushing or conduit in an unused

conduit hub on the bottom of the 12 VDC Beacon 200 housing.

5. Route the wires in conduit or shielded cable from the detector head

through the selected conduit hub into the 12 VDC Beacon 200. See
Table 5 below for wire size and distance guidelines.

Unshielded twisted pair cable in conduit or shielded twisted pair cable
is recommended for all the direct connect detector heads. For the LEL
detector, pair the R & B wires and the W & G wires. Shielded cable or
wires in conduit are recommended for the 2-wire and 3-wire 4 - 20 mA
transmitters.

6. Connect the wires from the detector head to the appropriate detector/

transmitter terminals. The top detector terminal strip is for channel 1
and the bottom one is for channel 2. See the detector head instruction
manual for controller terminal connections.

CAUTION: Do not route power and detector head wiring through the same

conduit hub. The power wiring may disrupt the transmission of
the detector head’s to the 12 VDC Beacon 200.

+

CH2
OUT

+

CH1
OUT

+

Recording Device
#2, 1Kohm Max
Impedance

Recording Device
#1, 1Kohm Max
Impedance

+

Figure 9. Recorder Output Wiring