3 Product Components

3-1. Main unit and standard accessories

- 9 -

GX-8000





3

Product Components




3-1. Main unit and standard accessories


After opening the package, check the main unit and accessories.
If anything in the following list is not included, contact RIKEN KEIKI.















• AC powered charger: 1


• Gas sampling probe and

gas sampling hose: 1



• Shoulder strap: 1


• Operating

manual

• Product

warranty


























GX-8000
main unit

Li-ion battery unit

DANGER

About explosion-proof

• Do not modify or change the circuit or structure, etc.

• When measuring the oxygen concentration, do not measure anything but a mixture of air and

combustible gases or vapors and toxic gases.

• When using this gas monitor in a hazardous area, take the following countermeasures for

preventing dangers resulting from electrostatic charges.

(1) Wear anti-static clothes and conductive shoes (anti-static work shoes).
(2) For indoor use, use the gas monitor while standing on a conductive work floor (with a

leakage resistance of 10 MΩ or less).

• The battery units that can be connected are the BUL-8000(G) (certificate number TC19437) or

BUD-8000(G) (certificate number TC19438).

• The specifications of the gas monitor are as follows:

Pump circuit:

Allowable voltage of 4.95 V, allowable current of 1.12 A, and
allowable power of 1138 mW

Combustible gas sensor circuit: Allowable voltage of 4.95 V, allowable current of 0.834 A, and

allowable power of 853 mW

Buzzer circuit:

Allowable voltage of 4.95 V, allowable current of 0.431 A, and
allowable power of 441 mW

Main circuit:

Allowable voltage of 4.95 V, allowable current of 0.717 A, and
allowable power of 733 mW

Backup circuit:

3.0 VDC, 10 µA