beautypg.com

RKI Instruments GX-8000(PT0-098) User Manual

Page 30

background image

4 How to Use

4-8. Display/setting mode

GX-8000

- 30 -

(TYPES-A and E Only)


The models that can display combustible gas levels in two ways, "%LEL range" and "vol% range,"
automatically switch between these two displays according to the gas concentration or oxygen
concentration, from "%LEL range" to "vol% range" and vice versa. They also support manual selection of
measurement range.

(1) Press the DISPLAY switch and select the combustible gas measurement range setting from the

display/setting mode menu.

The following screens are displayed in turn on the gas monitor.


(2) Press the ENTER switch.

NOTE

If you do not want to make a change, press the DISPLAY switch to return to the display/setting mode
menu.


(3) Every time the ▲ or ▼ switch is pressed, the measurement range menus, AUTO RANGE (automatic

switching), ONLY VOL (fixed to the vol% range), and ONLY LEL (fixed to the %LEL range), are
displayed in turn.

Press the ▲ or ▼ switch to select a measurement range and press the ENTER switch.


(4) When END is displayed, the setting is completed.

The display/setting mode menu is displayed again.


(5) After completion, press the DISPLAY switch several times until it returns to the detection mode.