4 How to Use
4-8. Display/setting mode
GX-8000
- 28 -
4-8. Display/setting mode
This mode allows you to change various displays and settings.
Every time the DISPLAY switch is pressed, various screens are displayed in turn.
(* Operations are slightly different depending on the type.)
LCD
Detection Mode
Display and Setting Mode
Combustible Gas Measurement
Range Setting
Used to select a combustible gas
measurement range manually.
(TYPES-A and E only)
ENTER
Combustible Gas Measurement
Range Setting
⇒ P30
PEAK
Display the maximum
concentration (or minimum
concentration for oxygen)
detected during measurement
from power-on to the present.
STEL
Display the STEL value after
power-on.
(TYPES-A, B, C, and D only)
TWA
Display the TWA value after
power-on.
(TYPES-A, B, C, and D only)
Full Scale Display/Alarm
Setpoint Display/Alarm Test
Display the full scale and alarm
setpoint values and check the
alarm operations for the settings
displayed.
ENTER
Full Scale Display/Alarm Setpoint
Display/Alarm Test
⇒ P32