4-7. Air calibration mode

GX-8000

- 26 -

4-7. Air calibration mode



















Press the AIR switch.

Keep pressing the AIR switch.
Changes the display to Adj -
HOLD AIR.

When RELEASE is displayed,
release the AIR switch.
A 30-second countdown is
started on the LCD.
(TYPES-A and E only)

When the zero adjustment is
successfully completed, END
is displayed, and you return to
the detection mode.

WARNING

When air calibration is performed in the atmosphere, check the atmosphere for freshness before
beginning the calibration. If other gases exist, the adjustment cannot be performed properly, thus
leading to dangers when the gas leaks.

CAUTION

• Perform air calibration under pressure and temperature/humidity conditions close to those in the

operating environment and in fresh air.

• Perform air calibration after the reading is stabilized.
• If there is a sudden temperature change of 15°C or more between the storage and operation

locations, turn on the power of the gas monitor, let it leave for about 10 minutes in a similar
environment to the operation location, and perform air calibration in fresh air before using it.