LATCHING ALARM
Toggles the instrument alarm logic between self-resetting (default)
and latching alarms.
Note
In the event of an alarm, the latching alarms option requires
the user to press DISPLAY/(ADJ.) to reset the alarms, even
after the gas concentration has dropped below the alarm
level.
SET LUNCH
Toggles the Lunchbreak feature on and off. With Lunchbreak off
(default), the STL TWA RESUME screen does not show at start-up.
SET ID
Toggles the ID INPUT option on and off. With the option off (default)
the user can not change the USER and ID values (see Display
Functions).
SET TIME CAL.
Sets the interval between calibration alerts on the GX-82A. Default
setting is off, with a range of 1 - 9999 hours. Tip: hold AIR or
(SHIFT) to scroll rapidly through values.
SET LOG ALARM
Sets the value of Data Logger memory capacity alerts on the GX-82A.
Default setting is off, with a range of 1 - 100%. Tip: hold AIR or
(SHIFT) to scroll rapidly through values.