Model GX-82A Instruction Manual

062796

RKI Instruments, Inc.

37

3. “SET ALARM or

Prompts changes for LOW, TWA, STL, and CLG (ceiling) alarms
for H

2

S or CO.

SET SERIAL No.

Sets the serial number for use by the Data Logger.

1. At the “SERIAL No.” screen, the first character (left end) will

flash. Press AIR or SHIFT to scroll through 0 - 9, A - Z to select
the desired character.

2. Press ON/OFF/ENTER to enter that value and move to the next

character.

3. Repeat steps 1. and 2. until the serial number is displayed, then

press ON/OFF/ENTER to go to the END screen.

SET DATE / TIME

Sets the date and time for use by the Data Logger.

1. The SET DATE / TIME screen shows the current settings, with the

month flashing.

2. Press AIR or (SHIFT) to select the desired month.

3. Press ON/OFF/ENTER to enter the selection and go to the next

setting.

4. Repeat steps 2 and 3 for each value: day, year, hours, minutes.

5. When the desire date/time is entered, press ON/OFF/ENTER to go

to the END.

SET INT. TIME

Sets the interval between logging events for the Data Logger.
Minimum value 10 seconds, maximum value 300 seconds.