Model GX-82A Instruction Manual
062796
RKI Instruments, Inc.
29
3. Bias Current Discharge
The batteries continuously supply a small current to maintain the
toxics sensors, even when the instrument is off (see Sensor
Maintenance). This current drain is minimal, but will result in a
normal discharge of the batteries over a period of several weeks.
Note
If the batteries are fully discharged before replacement,
allow 1/2 hour for the toxic gas circuit to show a normal
response.
Sensor Maintenance
Electrochemical sensors (O
2
, H
2
S, CO) gradually deteriorate,
regardless of use, and require periodic replacement. Combustibles
sensor life is generally related to usage, but other factors may affect
duration.
The GX-82A sensors are easy to replace, but only the combustibles
sensor contains user-serviceable components. If a sensor requires
replacement, call RKI or your local distributor. All sensors are
warranted usable for one year from the date of shipment. Sensors that
fail within the warranty period will be replaced at no charge.
Combustibles Sensor
Replace the combustibles sensor or filaments when:
1. The combustibles circuit cannot be calibrated correctly.
2. The HC (%LEL) display does not show 0 immediately after the
start-up sequence, and it cannot be set to zero by the Demand Zero
command.