Model GX-82A Instruction Manual

28

Hayward, California (510) 441-5656

MAINTENANCE

Batteries

1. Check the battery voltage periodically by pressing the DISPLAY

switch to reach the Battery Voltage function. Replace the batteries
before the voltage drops to the operational limit (see Operation-
Alarms).

WARNING

T

AKE

THE

GX-82A

TO

A

NON

-

HAZARDOUS

LOCATION

BEFORE

CHANGING

OR

CHARGING

THE

BATTERIES

.

2. To replace the batteries, remove the battery compartment cover.

Remove the batteries and verify that the battery compartment and
electrical contacts are clean. Insert fresh batteries (alkaline or
fully-charged Ni-Cad) according to the polarity (+/-) markings
and replace the cover.

3. Recharging Batteries

Note

The GX-82A charger detects the installed battery type and
will not charge alkaline cells, to prevent damage to the
instrument. Only Ni-Cad cells will charge properly with the
GX-82A charger.

Rechargeable batteries may be recharged inside the GX-82A in a
non-hazardous location. Turn the instrument off, connect the
charger to the charger jack, and plug the charger into a 115 V AC
outlet. Full charge requires approximately 12 hours.