Model GX-82A Instruction Manual
28
Hayward, California (510) 441-5656
MAINTENANCE
Batteries
1. Check the battery voltage periodically by pressing the DISPLAY
switch to reach the Battery Voltage function. Replace the batteries
before the voltage drops to the operational limit (see Operation-
Alarms).
WARNING
T
AKE
THE
GX-82A
TO
A
NON
-
HAZARDOUS
LOCATION
BEFORE
CHANGING
OR
CHARGING
THE
BATTERIES
.
2. To replace the batteries, remove the battery compartment cover.
Remove the batteries and verify that the battery compartment and
electrical contacts are clean. Insert fresh batteries (alkaline or
fully-charged Ni-Cad) according to the polarity (+/-) markings
and replace the cover.
3. Recharging Batteries
Note
The GX-82A charger detects the installed battery type and
will not charge alkaline cells, to prevent damage to the
instrument. Only Ni-Cad cells will charge properly with the
GX-82A charger.
Rechargeable batteries may be recharged inside the GX-82A in a
non-hazardous location. Turn the instrument off, connect the
charger to the charger jack, and plug the charger into a 115 V AC
outlet. Full charge requires approximately 12 hours.