Model GX-82A Instruction Manual

062796

RKI Instruments, Inc.

33

sensor cover, then slide the pump into place. Make sure the pump
housing seats and clips securely.

2. Turn on the pump and the GX-82A and allow it to complete the

start-up procedure.

3. Take the instrument into the area to be monitored.

Caution

Do not place the probe tip in liquid or other materials
which could be drawn into the Pump or the GX-82A.

4. Place the probe tip in the desired location to sample the atmo-

sphere. Watch the GX-82A’s display for changes. If the sample
system becomes blocked, the low flow alarm will sound.

WARNING

W

HEN

THE

S

AMPLE

-D

RAWING

P

UMP

IS

ATTACHED

,

IT

MUST

BE

RUNNING

FOR

THE

GX-82A

TO

DETECT

CHANGES

IN

GAS

CONCENTRATIONS

. T

HE

GX-82A

MAY

RESPOND

MORE

SLOWLY

TO

CHANGES

IN

THE

AMBIENT

AIR

. U

SE

THE

PUMP

ONLY

WHEN

SAMPLE

-

DRAWING

MEA-

SUREMENTS

ARE

NECESSARY

,

AND

REMOVE

IT

BEFORE

RESUMING

REGULAR

MONITORING

.

5. The Pump will run about 10 hours on a set of fresh “AA” batter-

ies. If the battery voltage drops too low, the low flow alarm will
sound. Take the GX-82A and Pump to a non-hazardous area
before replacing the batteries.

Manually-Operated Sample-Drawing Aspirator

The Sample-Drawing Aspirator accessory includes an adapter, hose,
aspirator bulb, and probe. It provides a low-cost solution to occasional