beautypg.com

RKI Instruments GX-82A User Manual

Page 35

background image

Model GX-82A Instruction Manual

32

Hayward, California (510) 441-5656

ACCESSORIES

Extender Cable

The Extender Cable extends the sensor block up to 10 meters from the
instrument, allowing confined space and remote monitoring, with all
the features of the GX-82A. The Extender cable includes the cable
with sensor block plug and socket, and a rubber guard for the sensor
cover.

1. Installing Extender Cable

Take the GX-82A to a non-hazardous area and turn off.

2. Remove the sensor cover.

3. Flip open the spring clips at the sides of the sensor block.

4. Pull the block straight up to unplug.

5. Plug the extender cable into the GX-82A, in place of the sensor

block. Flip closed the spring clips.

6. Plug the sensor block into the end of the Extender Cable. Close the

spring clips.

7. Slide the sensor cover over the block on the end of the Extender

Cable. Slip on the rubber guard.

8. Start-up the GX-82A and verify operation as normal.

Sample-Drawing Pump

The Sample-Drawing Pump accessory includes the pump, hose, and
probe. It provides continuous sample-drawn monitoring for a variety
of applications.

1. Turn off the GX-82A and take to a non-hazardous area.

2. Assemble the pump, probe, and hose. Remove the GX-82A’s