Model GX-82A Instruction Manual

10

Hayward, California (510) 441-5656

2. Verify Operation

To easily verify correct operation of the GX-82A, breathe out over
the diffusion grill of the instrument. The O2 reading should drop
measurably before returning to normal.

To verify detection of combustibles use a controlled source of
flammable vapor, for example, a bottle of isopropyl alcohol. The
audible alarm will sound, the %LEL will blink, and the HC will
flash on the display.

Caution

Do not use gas from a cigarette lighter to test response to
combustibles. Exposing the sensor to uncontrolled high
concentrations of gas in this manner will reduce
response and sensor life.

WARNING

I

F

THE

GX-82A

DOES

NOT

RESPOND

TO

THESE

VERIFICA-

TIONS

,

TAKE

IT

TO

A

KNOWN

FRESH

-

AIR

ENVIRON-

MENT

,

THEN

FOLLOW

THE

DEMAND

ZERO

PROCEDURE

DESCRIBED

IN

A

DJUSTMENT

AND

C

ALIBRATION

. R

EPEAT

THE

V

ERIFY

O

PERATION

PROCEDURE

BEFORE

USING

THE

GX-82A

IN

A

POTENTIALLY

HAZARDOUS

LOCATION

.

Normal Operation

The GX-82A will continuously monitor the atmosphere, and display
the LEL, O

2

, and toxic gas concentrations present. If the GX-82A is

taken into a low-light environment, the display backlight will
automatically turn on. To use with sample-drawing attachments, insert
the probe into the area to be monitored and wait a few seconds for