Servicing, 1 grinding discs, 2 replacing the grinding discs (5 + 6) – Retsch DM 400 User Manual
Page 42
Servicing
42
Function part
Task
Test
Service interval
Grinding chamber sen-
sor
Start inhibitor
Open grinding cham-
ber; start; error mes-
sage
Always before use
Grinding chamber safe-
ty interlock
Start inhibitor
Do not close grinding
chamber electrically;
start; error message
Always before use
Grinding container
protective equipment
Start inhibitor
Sample material con-
tainer not inserted;
start; error message
Always before use
Grinding disc gap width
Crush function
Check distance
Always before use
7.1 Grinding discs
The grinding discs wear after a certain period of time according to
sample material. Check the surfaces at specific intervals. Refer to
the information in the chapter
“Reverse operation” and the chapter
below “Replacing the grinding discs (5 + 6)” to change the direction of
rotation or to replace the discs.
7.2 Replacing the grinding discs (5 + 6)
CAUTION!
The device must be connected to the mains to open the
grinding chamber and to adjust the gap width.