New Era NE-4000 Multi-Phaser Programmable Double Syringe User Manual
Page 44

New Era Pump Systems Inc.
www.SyringePump.com
Model NE-4000 Multi-Phaser™
Publication #1200-01
39
09/05/14
BUZZER
BUZ [ 0 | { 1 [ < n > ] } ]
Sets / queries buzzer
Set:
0 = Turn buzzer off;
1 = Turn buzzer on
if
If
otherwise buzzer beeps
if
Query response: { 0 | 1 }
0 = Buzzer off
1 = Buzzer is on continuously or beeping.
9.4.5
System Commands
SET PUMP NETWORK ADDRESS AND BAUD RATE
* ADR [
[ B { 19200 | 9600 | 2400 | 1200 | 300 } ] ][DUAL | RECP | ALTR]
Set/query pump network address
=>Valid range: 0 to 99
B { 19200 | 9600 | 2400 | 1200 | 300 } will change the baud rate as indicated. NOTE: The command
response and all further communications will be at the specified baud rate.
Special communications modes
DUAL Sets the pump to Dual Pumping mode with a secondary pump.
RECP Sets the pump to Reciprocating Pumping mode with a secondary pump.
ALTR Sets Alternating pumping mode.
All special communications modes use 19,200 baud rate.
This is a special system command that will be accepted by the pump regardless of its current address or mode.
Once set, the pump will only respond to commands with the set address and at the specified baud rate.
Note: Once a special communications mode is set, the pump will only respond to commands that are preceded by
the ‘*’ character. To exit special communication mode, reset the address: * ADR 0
Example: *ADR
Query current address setting
*ADR 3
Set pump network address to 3. The pump will now only respond to
commands with address 3
*ADR 5 B 1200
Set the pump network address to 5 and the baud rate to 1200. The command
response will be at 1200 baud
ENABLE SAFE COMMUNICATIONS MODE
SAF [
Set/query Safe communications mode setting.
If
If
communications must be received every