Using the controls on the MTS Landmark load frame, slowly clamp the specimen in the upper grip.
2. If necessary, recover from a tripped limit (otherwise proceed to the next step).
Clamping can result in tripping a force or displacement limit. See
If Necessary, Recover from a Tripped
(p. 110) or
If Necessary, Recover from a Tripped Displacement Limit
(p. 113).
Check and Reset Meter History
Checking meter history after clamping the specimen in the upper grip allows you to verify that excessive
force was not applied to the specimen during clamping.
1. Check meter history.
Note: Acceptable limits are dependent on the specimen you are using for the test. For detailed
information, consult your lab experts or a materials handbook.
2. Reset meter history.
Now that specimen installation is complete, reset meter history by navigating to Station Manager >
Display > Meters and clicking the Reset button.
MTS Landmark Test System Operation | 85
Running the Example Tension Test