Note: For this test, MTS has determined that a setting of +5 kN will prevent damage to the
system and the specimen.

Description

Item

Type here and press Enter

1

Note: You must press Enter after typing numerical entries for them to be accepted.

a) Select Station Manager > Display > Station Setup > Channels > Ch1 (or other depending on

your configuration) > Force > Limits, click the Channel Input Signals button, and then click the
Limits tab.

b) Set the Upper Limit to +5 kN (a positive number indicates tension). Be sure to press Enter after

typing in the limit.

c) Then, in the Upper Action drop-down list, select Interlock.

3. Set the Force Lower (Compression) Limit.

MTS Landmark Test System Operation | 73

Running the Example Tension Test