MTS Series LX Laser Extensometer User Manual
Page 18
LX Laser Extensometer
Specifications
Introduction
18
Scan Line Orientation
Vertical
Scan Rate
100 scans/second
Unit System
Selectable, millimeter or inch
Target Distance
LX 500
LX 1500
Selectable: 304 mm (12 in), 381 mm (15 in)
Selectable: 381 mm (15 in), 457 mm (18 in)
Measurement Range
†
LX 500
LX 1500
5 to 127 mm (0.2 to 5.0 in)
8 to 381 mm (0.3 to 15 in)
Resolution
LX 500
LX 1500
0.001 mm (0.0001 in)
0.01 mm (0.001 in)
Nonlinearity (maximum)
‡
LX 500
LX 1500
±0.025 mm (±0.0001 in)
±0.15 mm (±0.006 in)
Repeatability
‡
LX 500
LX 1500
±0.003 mm (±0.0001 in)
±0.03 mm (±0.001 in)
Zero Suppression
Reading can be zeroed anywhere within the full scale range of the
instrument
Multiple Scan Averaging
A moving window average is applied over a programmable number
of scans to digitally filter the output readings
Display
2-line, 16 characters/line LCD type display with LED backlighting
Analog Output Range
±10 V full scale (±5 V full scale optional)
Serial I/0 Port
Standard 3-wire type, RS-232
Selectable baud (9600, 4800, 2400, 1200)
Options
Consult MTS
*
Specifications apply at 25
°
C (77
°
F), 50% relative humidity.
† Contact factory for other ranges.
‡ Over optimum displacement range at calibrated target distance.
P
ARAMETER
T
YPICAL
S
PECIFICATION*