Leica led5000 mci – Leica Biosystems M205A User Manual
Page 43
Leica M series
Manual
43
The Leica LED5000 MCI™ (for "Multi Contrast
Illumination") is installed using two screws. For
optimum accessibility, the optics carrier should
be removed during installation (see page 15.)
Constraints
The Leica LED5000 MCI™ cannot be used
together with the objective nosepiece.
Assembly
•
1. Hold the LED5000 MCI™ with one
hand and tightly screw the retraining
stirrups on both top holes on the drive hous-
ing.
2. Connect the CAN-bus cable to either of the
two sockets. (The flat part of the plug must
be facing downwards.)
3. Plug the other end of the cable into one of
the two "CTL2" sockets on the column.
Leica LED5000 MCI™