IsoPro™ Manual Cross-stage: Assembly (continued)
3. Screw the cross-stage evenly onto the three
threaded holes.
4. Push the cross-stage all the way back
towards the column.
5. Reinsert the glass plate back into the stan-
dard stage.
Focusing drive and column
1. Unscrew the extension plate from the base
using the Allen key provided.
2. Attach your focusing drive column to the
bottom using the three Allen screws.
3. Reattach the adapter plate to its original
position using the three Allen screws.