IsoPro™ Manual Cross-stage: Assembly
Before the IsoPro™ cross-stage is mounted to
the base, the axis containing the control buttons
is attached either on the left or the right side of
the cross-stage.
IsoPro™ Cross-stage
Before the IsoPro™ cross-stage is mounted to
the base, the axis containing the control buttons
is attached either on the left or the right side of
the cross-stage.
If the controls are to be mounted on the left-
hand side, the gear rod on the bottom side of
the cross-stage must be unscrewed and reat-
tached in reverse.
1. Take the glass plate from the cross-stage.
2. Turn the cross-stage around and place it
onto a non-slip surface.
3. Change the gear rod from the left to the
right-hand side.
4. Skip the next two steps to mount the
controls.
Left or right operation
If the controls are to be mounted on the left-
hand side, the gear rod on the bottom side of
the cross-stage must be unscrewed and reat-
tached in reverse.