Troubleshooting – Leica Biosystems ST4020 User Manual
Page 42

42
Instructions for Use V1.1, RevB – 10/2012
•
Keypad does not respond to
key presses
•
The keypad or the control
board is faulty.
•
Inform Technical Service.
Possible cause
Issue
Corrective action
•
Water does not flow.
•
Water hose is not
connected.
•
Tap is not open.
•
The valve in the instrument
is faulty (does not open) or
the control board is faulty.
•
Adjust the flow rate for the
rinsing water at the tap as
described in Chap. 5.4.
•
The drain hose must not be
kinked.
•
Check the drain hose for
objects that might block it.
•
Place the drain hose so that
water does not have to rise
above the level of the drain
fitting.
•
Water overflows at the rin-
se stations.
•
Tap is open too far.
•
The drain hose is clogged.
•
The drain hose is routed
incorrectly, gradient
insufficient.
7.
Troubleshooting
•
Connect the water hose to
the stainer and the tap.
•
Turn on the tap.
•
Inform Technical Service.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)