Optional accessories – Leica Biosystems CM1950 User Manual
Page 63

63
Leica CM1950 – Cryostat
Cleaning the knife holder CN
Daily cleaning
For daily cleaning, it is enough to fold the anti-roll
guide system forwards and remove the section
waste from the knife holder using a dry brush.
Please use a cold brush, as otherwise the sec-
tion waste will thaw and stick to the knife holder.
Fig. 72
1. Fold the anti-roll guide system (33) to the left
while holding it down by the lever (34).
2. Unscrew the bolts (35) of the pressure plate.
3. Afterwards, the pressure plate (36) can be re-
moved for cleaning (with alcohol or acetone).
11.
Optional Accessories
Disinfection (e.g. with Leica Cryofect)
Spray contaminated surfaces with a uniform coat of concentrate or wipe
them down with a rag soaked in it, allow it to make contact for 15 minutes,
then wipe it off.
33
34
35
35
36
For disinfection, commercially avail-
able, mild cleaners and disinfectants
can be used – we recommend Leica
Cryofect.
It is not necessary to oil the parts, such
as the T-piece on the microtome base
plate, clamping lever etc.
The safety gloves included in the stan-
dard scope of delivery must be worn
when cleaning the blade holder to
prevent frostbite of the skin.
Cleaning the blade holder CE
Daily cleaning
The use of flammable sprays within the cryostat chamber is permit-
ted only with the Leica Cryofect tested by us.
Fig. 71