Safety – Leica Biosystems LPC Operating Manual User Manual
Page 9

9
Leica LPC, Laser Cassette Printer
2.
Safety
Warnings – Electrical safety
• Connect only the Filtronic GD80 filter unit provided with your LP C cassette to the exhaust
output (100-240 V, 50/60 Hz, 1 A max.).
• The supply cable of the Leica LP C must be connected to a grounded socket only. Ensure that
the Leica LPC Laser Cassette Printer is installed such that the power supply cable on the rear
side of the instrument can be removed easily.
• The supply cable of the Leica LP C must be connected to a grounded socket only. Use only
the power supply cable provided.
Warnings – Mechanical safety
• Never attempt to operate the Leica LP C Laser Cassette Printer without the protective cover
in place.
• Never attempt to bypass any of the safety devices on the instrument.
• In addition to the Class 4 Laser, the instrument contains moving machine parts that can injure
the user if the instructions above instructions are not observed (see Chap. 7 - Appendix A on
page 38).
• Never use the Leica LP C Laser Cassette Printer without ensuring that the filter unit is con-
nected properly and is working. Doing so will cause the instrument to malfunction and expel
harmful dust particles.
• Always set up the Leica LP C on a stable, flat surface. Set up the LP C Laser Cassette Printer
so that the front and rear operating elements are easily accessible to the operator.
• Remember that minor color changes can occur in LP C cassettes if these are exposed to re-
agents that contain Picric acid, e.g. Bouin's fixative.
• Remember that text imprinted on Leica LP C cassettes can become illegible or be lost if these
are exposed for long periods to solutions that contain strong acids. Decalcification process-
es of tissue in Leica LP C cassettes that involve strong acids such as nitric acids, must be
checked by the users before implementation.