Leica Biosystems LPC Operating Manual User Manual
Page 34

34
Operating Manual V 1.0 – 06/2010
•
Remove the filter cartridge (22) by reaching
into the opening with both hands and pulling it
upwards and out.
Fig. 32
•
Replace the old main filter with a new one and
mark the top of the filter with the date for the
next replacement.
IMPORTANT: So that the main filter is
installed correctly, point the blower
inlet to the main filter outlet that is lo-
cated at the bottom of the main filter.
Fig. 34
22
5.
Cleaning and Maintenance
Bottom of
the main
filter
When the main filter is installed the top
of the filter should be marked by the
user with the recommended replace-
ment date. (Is visible when the cover of
the main filter unit is removed.)
Fig. 33