4.5
Installation site requirements
The installation site must be well-ventilated.
Never operate the instrument in rooms with explosion hazard.
4.
Instrument Setup
•
The printer needs an installation area of approx. 260 x 460 mm and the fil-
ter unit needs an additional installation area of approx. 470 x 490 mm.
Additional space is required for the PC containing the control software.
•
Stable laboratory bench and practically vibration-free floor.
•
Room temperature consistently between +5 °C and +40 °C.
•
Relative air humidity maximum 80 %, non-condensing.
•
Avoid vibrations, bright direct light and strong temperature fluctuations.
•
The AC socket used for the power supply must be close to the laser
printer and easily accessible.
•
The laser printer is intended for use in enclosed rooms only and may be
used at elevations up to 2000 m only.