Function instructions – Klein Tools DTL302 User Manual
Page 7
7
• Press “ ” or “
” until the display shows “
”.
• Press “ENTER” to display “ ” or “
” in the time display.
If “
” is displayed, press “ ” or “
” to select “ ”.
• Press “ENTER” to display choice number 3 in the secondary
display and
00:30 h:m
in the time display.
• Press “ ” or “
” until the display shows the Auto-Power-
Off Time choice number you want, and then press “ENTER” to
store the Auto-Power-Off setting in memory.
• Auto-Power-Off mode is automatically disabled in MIN/MAX
and logging modes.
H.
Change Logging Interval
• The logging interval determines how often the thermometer
stores logged readings in memory. You choose the length of
the logging interval. The thermometer stores logged readings
at the end of each logging interval.
• Select logging intervals
• Press “ ” or “
” until the display shows
and “ LOG ”.
• Press “ENTER” to display the logging interval choices.
• Press “ ” or “
” until the display shows the logging
interval you want, and then press “ENTER” to select.
• For user-defined logging interval (1 sec. to 23 hrs : 59 mins).
Press “ ” or “
” until the display shows “h:m” or “m:s”,
and then press “ENTER” to select. The last number in the
time
display
blinks.
Press “ ” or “
” until the logging interval you want
appears on the display, and then press “ENTER” to select.
• Holding down “ ” or “
” causes the number to change
more quickly.
I.
Setting Temperature Offset
• You can adjust the thermometer’s readings to compensate
for the errors of a specific thermocouple. The allowable
compensation range is ±5.0°C or K, and ±9.0°F. The DTL302
allows you to store individual offsets for T1 and T2.
• Press “ ” or “
” until the display shows “ OFFSET ”.
• Press “ENTER” to display T1 reading (blinking), T1 offset.
• Press “ ” or “
” to display T1 or T2 in the primary display
to select 1 input.
• Press “ ” or “
” until the primary display shows the
correct reading, and then press “ENTER” to store the offset
setting in memory.
• NOTE: Reset the offset to 0.0 when it is no longer needed.
• NOTE: If T1 or T2 has no offset, “ OFFSET ”will not be displayed.
J.
Resetting The Thermometer
• To restore to factory settings, press “ ” or “
” until
the display shows “
” then press “ENTER” to display
“
“ in the primary display. Press “EXIT” to exit setup.
FUNCTION INSTRUCTIONS
1.
Connecting Thermocouples
• Thermocouples are color coded based on the ANSI color code:
• Plug a thermocouple into the input terminal(s).
• Set the thermometer for the correct thermocouple type. To
change the thermocouple type, see:
Set Up Instructions section,
Set Up Instructions
Set Up Instructions
F: Thermocouple Type
1.
:01 sec
2.
:10 sec
3.
1:00 min
4.
10:00 min
User Defined
J Type = Black
K Type = Yellow
T Type = Blue
E Type = Purple