Argument errors
Argument errors are displayed when a command argument
is unrecognized or out of range. The argument in question
will be displayed except in the case of missing arguments.
The argument error has the following format:
System commands
ADDRESS
Sets or displays the pump address. Valid range is 0 to 99.
Command format:
address [0-99]
Query response:
CATALOG
Displays a catalog of programs stored in the pump.
Command format:
cat
Query response:
This line is repeated for each file on the disk
The last line shows the number of files and the total number
of steps being used on the disk.
✐
NOTE: Pumps with an address of 0 are
masters, and pumps with an address
between 1 and 99 are slaves.
✐
NOTE: The file size is the number of steps in
the program plus one for the file information
header.
CONFIG
Sets or displays pump configuration.
Command format:
Config [{config designation}|{value}]
DELMETHOD
Deletes a program, stored in the pump.
Command format:
delmethod {program}
DIM
Sets or displays the brightness level in percent. Valid range
is 0 to 100. Note that specifying 0 will turn the backlight
off so that the screen will not be visible.
Command format:
dim [0-100]
Query response:
ECHO
Sets or displays the USB echo state. Valid states are on
or off.
Command format:
echo [on|off]
Query response:
or:
FORCE
Sets or displays the infusion force level in percent. Valid
range is 1 to 100.
Command format:
force [1-100]
Query response:
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
✐
NOTE: After changing the configuration, the
pump must be power cycled.