KD Scientific Legato 100 Series Syringe Pump User’s Manual
Publication 5617-006 Rev D
5. Screw down the syringe clamp knob (6) until the
syringe barrel clamp (10) is tight against the
syringe barrel. Do not overtighten, especially
when using glass syringes.
6. Secure the syringe flanges by pushing the retaining
brackets (4,5) tightly against the syringe flanges
using the bracket clamping knobs (7,8). The
syringe plunger should be secured by the pusher
block retaining bracket and the flange of the
syringe should be secured by the syringe holder
block retaining bracket.
7.
Tighten the mechanical stop collars (9) on the
pump to prevent excess pusher block travel.
✐
NOTE:
The syringe barrel clamp (10) [single
only] may need to be flipped depending on
your syringe size. To do so, completely
unscrew and remove the syringe barrel
clamp knob from the syringe block. Then
remove lock nut (11). Flip the bracket and
reassemble.
SYRINGE LOADING
2
5
4
1
3
9
7
8
10
6
> 30 cc syringes
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
✐
NOTE: For infuse/withdraw models, make sure
the plunger is secured in the retaining bracket
and the barrel of the syringe is secured in the
retaining bracket.
The Legato 100 Series can accommodate up to two syringes.
Depending on the mechanism, any type of syringe, glass,
plastic or stainless steel can be used. The pump can
accommodate from 0.5 µl to 60 ml single syringe and 0.5 µl
to 10 ml dual syringe. The syringe barrel clamp should be
flipped for syringes smaller than 30 ml. For syringes larger
than 30 ml, the syringe clamp should be used as shown.
1.
Squeeze the release button (1) and move the
pusher block (2) to the approximate length of the
syringe.
2. Unscrew the bracket clamping knobs (7,8) to
loosen the retaining brackets (4,5) on the pusher
block (2) and the syringe holder block (3).
3. Unscrew the syringe barrel clamp (10) by
unscrewing the syringe barrel clamp knob (6).
4. Place the syringe on the syringe holder block (3).
Ensure the barrel flange is within the syringe holder’s
block’s retaining bracket (5) and the plunger flange is
in the pusher block’s retaining bracket (4).
< 30 cc syringes
11
!
CAUTION:
Be sure mechanical stop
collars are positioned properly to prevent
the plunger from bottoming out on glass
syringes.
✐
NOTE:
Syringe brackets (4 & 5) are only
available on Legato 110, 111 & 180.
10
Dual Syringe
Clamp