Setting the Force Limits
The maximum linear force of the syringe pump is 30 lbs
or 13.6 kgs. This force is sufficient to damage delicate,
low volume syringes, as well as many standard syringes if
pumping viscous fluids. The % of maximum force applied
by the pusher block can be changed.
In the set up screen, press
Force
.
Set the force level.
Press
Other Value
to enter a custom force %
Accept
the changes made by pressing
and go back to the Set Force screen.
Cancel
the changes made by pressing
and go back to the Set Force screen.
For a Custom Force %
Enter the precentage for custom force %
Accept
the changes made by pressing
and go back to the Set Force screen.
Cancel
the changes made by pressing
and go back to the Set Force screen.
✐
NOTE: A minimum value of approximately
30% is required to overcome friction forces
in the mechanism. You may have to set the
force level to a higher value depending on
the backpressure/friction in your system.
Press Here
Press here to enter
custom force %
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
Suggested Force Level Settings for Common Syringes
Syringe Type/Material
Capacity
Force Setting
Common Manufacturer(s)
Plastic Syringes
≤5 ml
50%
BD plastic, Sherwood,
Airtite, Terumo
>5 ml
100%
Stainless Steel
All
100%
KDS, Cole-Parmer
Glass/Glass
≤20 ml
30%
KDS Glass, Cadence (Popper)
>20 ml
50%
Glass/Plastic
≤1000µl
30%
Hamilton, SGE
≤5 ml
50%
>5 ml
100%