Kds model 200 series, Specifications – KD Scientific 200 Series User Manual
Page 5

1
KDS Model 200 Series
Specifications
Model
200 Series
Syringe size
10 microliter to 140 milliliter
Electrical Rating
US model 115 V~, 0.25 A
CE model 230 V~, 0.16 A
Fuse
5 x 20 mm, 250 V~, Slow Blow, 0.25 A
Voltage operating range
US model 110-120 V~, 50/60 Hz
CE model 200-240 V~, 50/60 Hz
Drive mechanism
Microprocessor controlled stepper motor
1/2 - 1/16 microstepping, driving a leadscrew
through a belt and pulley drive mechanism
Pusher advance per microstep
(1/16 step) - 0.165 micron (or .0000064 inch)
Volume per microstep
(1/16) with 60ml BD syringe - 0.0919µl
Max stepping rate
1600 (1/2 step) /sec
Min step rate
1 (1/16 step) /120 secs
Linear Travel Rate
Min: 4.95 x10
-4
cm/hr
Max: 12.67 cm/min
Flowrate range
2.757 µl/hr to 70.56 ml/min (60 ml syringe)
Nominal linear force
35 lb
Dimensions
11x9x5.5 inch (28x23.5x14 cm)
Weight
9 lb (4 kilo)
Atmospheric Specifications
Temperature
5˚C - 40˚C (41˚F - 104˚F)
Humidity
20% - 80% RH
Mode of Operation
Continuous
Classification
Class I
Pollution Degree
2
Installation Category
II
Output
N/A
Physiological Effects
N/A
Cooling Conditions
No special considerations
Mechanical Stability
No special considerations
Protective Packaging
No special considerations
Earth Terminals
No External connections required
Removable Protective Means
N/A
Supplier Name
KD Scientific Inc.
Address
84 October Hill Road, Holliston, MA 01746-1388