beautypg.com

Kds model 200 series, Standard minimum and maximum flow rates – KD Scientific 200 Series User Manual

Page 36

background image

32

KDS Model 200 Series

Standard Minimum and Maximum Flow Rates

Syringe size

Diameter*

Minimum

Maximum

10 µl

0.46 mm

0.001 µl/hr

21.10 µl/min

25 µl

0.73 mm

0.003 µl/hr

53.15 µl/min

50 µl

1.03 mm

0.005 µl/hr

105.8 µl/min

100 µl

1.46 mm

0.009 µl/hr

212.6 µl/min

250 µl

2.3 mm

0.021 µl/hr

527.6 µl/min

500 µl

3.26 mm

0.042 µl/hr

1060 µl/min

1 ml

4.61 mm

0.083 µl/hr

2119 µl/min

2.5 ml

7.28 mm

0.207 µl/hr

5286 µl/min

3 ml

8.59 mm

0.288 µl/hr

7360 µl/min

5 ml

10.3 mm

0.414 µl/hr

634 ml/hr

10 ml

14.57 mm

0.828 µl/hr

1270 ml/hr

20 ml

19.05 mm

1.414 µl/hr

2171 ml/hr

30 ml

21.59 mm

1.817 µl/hr

2789 ml/hr

50 ml

28.9 mm

3.277 µl/hr

4998 ml/hr

60 ml

26.6 mm

2.757 µl/hr

4234 ml/hr

100 ml

34.9 mm

4.746 µl/hr

7289 ml/hr

140 ml

38.4 mm

5.746 µl/hr

8824 ml/hr

Syringes from different manufacturers can have slightly different limits.

Note: *This is a reference diameter used to calculate the flow rate. The specific diameter

should be entered for your syringe type.