13

KDS Model 200 Series

Operating Instructions

(Continued)

CLEARING A STALL CONDITION

Should a stall occur the pump motor is stopped to prevent damage.

To clear the display press select.

To move the stalled mechanism use the fast forward or fast reverse to move the pusher

block. Using the fast forward or fast reverse feature is not only the most simple way to deal

with the stall, it also reduces potential damage to the cam mechanism which releases the

halfnut from the leadscrew.

NV Ram Failure

If the settings in the non-volatile memory become corrupted the display will read "NV Ram

Failure" and the pump will not operate.

To recover from this condition the pump must be powered down and then turned on again

after a few seconds delay. The pump will be re-initialized to the default settings and can now

be programmed as normal.

If the above fails to work the “NV Ram Failure” message can also be cleared by pressing

select and programming a new flow rate. The pump should then be turned off and on to

save the settings.

RS232 Setup

The RS232 connections are made through two modular telephone connectors, labeled IN

and OUT , located on the rear panel.

A single PC can control up to 100 pumps via a "daisy chain" using the IN and OUT

connectors. When using the daisy chain each pump must be assigned an address and set to

the same baud rate. A splitter may be required with greater than 50 pumps.

When controlled via RS232 the pump will still respond to keypad commands but will not

respond to keypad and RS232 commands simultaneously. All RS232 command settings,

similar to keypad settings, are stored in non-volatile memory.

Select RS232 from the main menu.

Baud Rate

300, 1200, 2400, 4800, 9600

The available baud rates will now be displayed and can be selected

using the arrow and select keys.

The display now prompts for assignment of a pump address.

Address

If, no address is assigned then the pump defaults to a 0 setting.

All pumps with the same address respond simultaneously to the same

commands.

Use the

key to enter an address, 0 - 99 and press select to return

to the main menu.