beautypg.com

Kds model 200 series, Programmable option– running the program – KD Scientific 200 Series User Manual

Page 26

background image

22

KDS Model 200 Series

KDS Model 200 Series

Programmable Option– Running The Program

Run

The Run key starts the program; the displayed time counts down and the direction

arrow flashes.

Hold/Continue

If the Run/Stop key is pressed while running a program, the pump is stopped but gives

an option to end the program, or restart the pump and continue the program to its end.

Program changes when operating

Once a step has commenced no changes are possible in that step. However, while

dispensing changes are permitted to steps still to be executed.

Syringe size changes

If the pump was previously used in Program Mode the pump will be initialized in

Program Mode when it is switched on. For convenience, it is possible to enter a new

syringe size, either from the stored Table or DIA, without leaving the Program Mode.

Note: If a syringe change is made this will change all program values to the default

settings and will require reprogramming. A diameter change causes the pump to stop;

resets the "number" of steps to 1; resets the "activestep" to 1; and all values will be set

to the initial default settings. The initialization of the new settings takes approximately

two seconds.

Stall Condition

The Fast Forward & Fast Reverse features do not work in Program Mode. Should a stall

occur then go to Infusion Mode where the Fast Forward/Fast Reverse features works,

and use these features to end the stall condition.

By going to the infusion mode the program is still saved in memory.

Changing programs containing loops

Only two loops are permitted in a program and it is necessary to remove from memory

the previously used programs which contain loops. A convenient way to do this is to

input a one step program which does not have a loop. This way, all loops are deleted

from memory and the new program containing loops can be added.