Kds model 200 series, Programmable option– rs232 commands and responses – KD Scientific 200 Series User Manual
Page 30

26
KDS Model 200 Series
Programmable Option– RS232 Commands and Responses
(Continued)
PROGRAM EXAMPLE
Syringe selected, 4.70 mm diameter
RS232 programming
mode prgm
Select Program mode
Number 4
Sets number of steps in the program
Step 1
Selects Step 1 for programming
time 00:00:10
Step 1 time duration is 10 seconds
travel I
Infusion selected
rateb 0 mlm
Step 1 beginning rate is 0 ml/minute
ratef 1 mlm
Step 1 finishing rate is 1 ml/minute
portout hh
Output pins 1 and 6 set at high/high
pause n
Pause inactive
loop n
No loops
save
Save step settings
Step 2
Ready to program step 2
time 00:00:15
Set time duration to 15 seconds
rateb 1 mlm
Assumes previous travel direction (infusion), and sets step 2 beginning
rate 1 ml/minute.
ratef 0.1 mlm
Finishing rate 0.1 ml/minute
loop y
Select a loop
loopto 1
Program will loop back to step 1 after completeing step 2
loopcnt 1
Will repeat the loop one time
save
step 3
Ready to program step 3
time 00:00:20
Time of step 3 is 20 seconds
rateb .3 mlm
Sets begin rate to 0.3 ml/min. Assumes no direction change.
ratef 0 mlm
Sets finish rate to 0 ml/min.
save
Step 4
Program step 4
time 00:00:12
Time duration 12 seconds
travel w
Change direction to withdrawal
rateb 1 mlm
Withdraw rate set to 1.0 ml/minute.
ratef 1 mlm
Finish rate 1 ml/min.
loop y
Select a loop
loopto 3
After step 4 will loop back to and repeat step 3.
loopcnt 1
Will repeat loop one time.
save
done
completes and saves program
Queries:
a) loops?
S2:1 S4:1
loop in step 2, loop count is 1; loop in step 4,
count 1
b) step 3 portout?
HH
Portout set in step 1 and remained unchanged
c) step 1 ratef?
1 ml/m
Finish rate in step 1 is 1.0 ml/minute