beautypg.com

Technical data – JUMO 701550 di 308 - Digital Indicator Data Sheet User Manual

Page 2

background image

2012-05-30/00485042

Data Sheet 701550

JUMO GmbH & Co. KG • 36035 Fulda, Germany

Page 2/7

Technical data

Thermocouple input

RTD input

Input for standard signals

Binary inputs

Measuring circuit monitoring

1

The accuracy refers to the max. measurement range span. The linearization accuracy is reduced with short spans.

Designation

Measuring range

Meas. accuracy

1

(incl. cold junction)

Ambient
temperature error

Fe-Con L
Fe-Con J
Cu-Con U
Cu-Con T
NiCr-Ni K
NiCr-Con E
NiCrSi-NiSi N
Pt10Rh-Pt S
Pt13Rh-Pt R
Pt30Rh-Pt6Rh B
W5Re-W26Re C
W3Re-W25Re D
W3Re-W26Re
Chromel-copel

EN 60584

EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584

GOST 8.585-2001

-200 to

+900°C

-200 to +1200°C
-200 to

+600°C

-200 to

+400°C

-200 to +1372°C
-200 to +1000°C
-100 to +1300°C

-50 to +1768°C
-50 to +1768°C

0 to +1820°C
0 to +2320°C
0 to +2495°C
0 to +2400°C

-200 to

+800°C

 0.25%
 0.25%
 0.25%
 0.25%
 0.25%
 0.25%
 0.25%
 0.25% (from 0°C)
 0.25% (from 0°C)
 0.25% (from 300°C)
 0.25%
 0.25%
 0.25%
 0.25%

100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C

Cold junction

Pt100, internal

Designation

Connection circuit

Measuring range

Meas. accuracy

1

Ambient
temperature error

3-/4-wire

2-wire

Pt100

EN 60751

2-wire / 3-wire / 4-wire

-200 to +850°C

 0.05%

 0.4%

50ppm/°C

Pt500

EN 60751

2-wire / 3-wire / 4-wire

-200 to +850°C

 0.2%

 0.4%

100ppm/°C

Pt1000

EN 60751

2-wire / 3-wire / 4-wire

-200 to +850°C

 0.1%

 0.2%

50ppm/°C

Pt50

GOST 6651-94

2-wire / 3-wire / 4-wire

-200 to +850°C

 0.1%

 0.8%

50ppm/°C

Pt100

GOST 6651-94

2-wire / 3-wire / 4-wire

-200 to +850°C

 0.05%

 0.4%

50ppm/°C

Cu50

GOST 6651-94

2-wire / 3-wire / 4-wire

-50 to +200°C

 0.2%

 1.6%

50ppm/°C

Cu100

GOST 6651-94

2-wire / 3-wire / 4-wire

-50 to +200°C

 0.1%

 0.8%

50ppm/°C

KTY11-6

2-wire

-50 to +150°C

 2.0%

50ppm/°C

Sensor lead resistance

30

 max. per lead for 3-wire/4-wire circuit

Measuring current

approx. 250µA

Lead compensation

Not required for 3-wire or 4-wire circuit. With a 2-wire circuit, the lead resistance can be compensated
in software by a correction of the process value.

Designation

Measuring range

Meas. accuracy

1

Ambient
temperature error

Voltage

0(2)—10V
0—1V
Input resistance R

IN

> 100k

 0.05%
 0.05%

100ppm/°C
100ppm/°C

Current

0(4)—20mA, voltage drop

1.5V

 0.05%

100ppm/°C

Resistance transmitter

min. 100

, max. 4k

4

100ppm/°C

Floating contacts

open = not active; short-circuit to GND = active

Transducer

Detection of

overrange/underrange

Detection of

probe/lead short-circuit

Detection of

probe/lead break

Thermocouple

yes

no

yes

RTD

yes

yes

yes

Voltage

2 — 10V
0 — 10V
0 — 1V

yes
yes
yes

yes

no
no

yes

no
no

Current

4 — 20mA
0 — 20mA

yes
yes

yes

no

yes

no

Resistance transmitter

no

no

yes

In the event of a fault, the outputs move to a defined (configurable) status.