Technical data – JUMO 701550 di 308 - Digital Indicator Data Sheet User Manual
Page 2

2012-05-30/00485042
Data Sheet 701550
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Page 2/7
Technical data
Thermocouple input
RTD input
Input for standard signals
Binary inputs
Measuring circuit monitoring
1
The accuracy refers to the max. measurement range span. The linearization accuracy is reduced with short spans.
Designation
Measuring range
Meas. accuracy
1
(incl. cold junction)
Ambient
temperature error
Fe-Con L
Fe-Con J
Cu-Con U
Cu-Con T
NiCr-Ni K
NiCr-Con E
NiCrSi-NiSi N
Pt10Rh-Pt S
Pt13Rh-Pt R
Pt30Rh-Pt6Rh B
W5Re-W26Re C
W3Re-W25Re D
W3Re-W26Re
Chromel-copel
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
GOST 8.585-2001
-200 to
+900°C
-200 to +1200°C
-200 to
+600°C
-200 to
+400°C
-200 to +1372°C
-200 to +1000°C
-100 to +1300°C
-50 to +1768°C
-50 to +1768°C
0 to +1820°C
0 to +2320°C
0 to +2495°C
0 to +2400°C
-200 to
+800°C
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25%
0.25% (from 0°C)
0.25% (from 0°C)
0.25% (from 300°C)
0.25%
0.25%
0.25%
0.25%
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
100ppm/°C
Cold junction
Pt100, internal
Designation
Connection circuit
Measuring range
Meas. accuracy
1
Ambient
temperature error
3-/4-wire
2-wire
Pt100
EN 60751
2-wire / 3-wire / 4-wire
-200 to +850°C
0.05%
0.4%
50ppm/°C
Pt500
EN 60751
2-wire / 3-wire / 4-wire
-200 to +850°C
0.2%
0.4%
100ppm/°C
Pt1000
EN 60751
2-wire / 3-wire / 4-wire
-200 to +850°C
0.1%
0.2%
50ppm/°C
Pt50
GOST 6651-94
2-wire / 3-wire / 4-wire
-200 to +850°C
0.1%
0.8%
50ppm/°C
Pt100
GOST 6651-94
2-wire / 3-wire / 4-wire
-200 to +850°C
0.05%
0.4%
50ppm/°C
Cu50
GOST 6651-94
2-wire / 3-wire / 4-wire
-50 to +200°C
0.2%
1.6%
50ppm/°C
Cu100
GOST 6651-94
2-wire / 3-wire / 4-wire
-50 to +200°C
0.1%
0.8%
50ppm/°C
KTY11-6
2-wire
-50 to +150°C
2.0%
50ppm/°C
Sensor lead resistance
30
max. per lead for 3-wire/4-wire circuit
Measuring current
approx. 250µA
Lead compensation
Not required for 3-wire or 4-wire circuit. With a 2-wire circuit, the lead resistance can be compensated
in software by a correction of the process value.
Designation
Measuring range
Meas. accuracy
1
Ambient
temperature error
Voltage
0(2)—10V
0—1V
Input resistance R
IN
> 100k
0.05%
0.05%
100ppm/°C
100ppm/°C
Current
0(4)—20mA, voltage drop
1.5V
0.05%
100ppm/°C
Resistance transmitter
min. 100
, max. 4k
4
100ppm/°C
Floating contacts
open = not active; short-circuit to GND = active
Transducer
Detection of
overrange/underrange
Detection of
probe/lead short-circuit
Detection of
probe/lead break
Thermocouple
yes
no
yes
RTD
yes
yes
yes
Voltage
2 — 10V
0 — 10V
0 — 1V
yes
yes
yes
yes
no
no
yes
no
no
Current
4 — 20mA
0 — 20mA
yes
yes
yes
no
yes
no
Resistance transmitter
no
no
yes
In the event of a fault, the outputs move to a defined (configurable) status.