beautypg.com

Troubleshooting guide / error messages – JULABO SC 10000w Eco-Series Recirculating Coolers User Manual

Page 39

background image

Recirculating Cooler

39

9.6. Reset

Use this to reset all values to factory setting.

Factory setting: no

1.

Press the

button until the menu item

is displayed.

2.

Press enter

to indicate the parameter (no).

3.

Press the keys

to select the parameter (YES).

4.

Press enter

to confirm the reset.

After the reset is carried out, the parameter is automatically set to
>no<.





10. Troubleshooting guide / Error messages

+

Alarm

with a complete shutdown of the unit

Whenever the microprocessor electronics registers a failure, a complete
shutdown of the heater and circulating pump is performed. The alarm light

“ illuminates and a continuous signal tone sounds. The MULTI-DISPLAY

(LED) indicates the cause for the alarm in form of a code.

Warning

without a complete shutdown of the unit

The MULTI-DISPLAY (LED) indicates the cause for the warning in form of a
code and an acoustic signal sounds in regular intervals.
These messages appear every 10 seconds.

Press enter

to quit the audible signal.

 The circulator is operated without bath fluid, or the liquid level is

insufficient. Replenish the bath tank with the bath fluid.

 Tube breakage has occured (insufficient filling level due to excessive bath

fluid pumped out). Replace the tubing and replenish the bath tank with the
bath fluid.

 The float is defect (e. g., because damaged in transit). Repair by

authorized JULABO service personnel.

 Excess temperature sensors defect.
 The excess temperature value lies below the working temperature

setpoint. Set the excess temperature to a higher value.