Technical specifications – JULABO SC 10000w Eco-Series Recirculating Coolers User Manual
Page 10

Operator responsibility – Safety instructions
10
2.4. Technical
specifications
Eco- Series
SC 5000a
SC 5000w
Working temperature ranges:
Standard unit
°C
+5 ... +35
+5 ... +35
Option Low Temp
°C
-20 ... +35
-20 ... +35
Option Low /HighTemp I
°C
-20 ... +80
-20 ... +80
Option Low /HighTemp II
°C
-20 ... +130
-20 ... +130
Temperature stability
°C
A
bsolute Temperature Calibration
0.1
0.1
±3
±3
Heater wattage (at 230 V) Option H5
kW
5.0
5.0
Cooling capacity
°C
+20 0 -10
+20 0 -10
Medium ethanol
kW
5.0 2.5 1.2
5.0 2.5 1.2
Refrigerant
R404A
R404A
Pump capacity P3 (Standard) / P4 * (Option)
see table 1 page 12
Flow rate
P3 / P4
l/min at 0 bar
34 / 42
34 / 42
Pressure max.
P3 / P4
bar at 0 liters
3.2 / 5.2
3.2 / 5.2
Overall dimensions (WxDxH)
cm
59x67/112
59x67/112
Filling volume
liters
43 ... 60
43 ... 60
Weight
kg 153
153
Ambient temperature
°C
5 ... 40
5 ... 40
Mains power connection
V/ Hz
365-440V/3PNPE/50Hz
400/50/3 phases
400/50/3 phases
Current consumption without heater
A
11/ phase (at 400 V)
11/ phase (at 400 V)
Current consumption with 5kW heater
A
23/ phase (at 400 V)
23/ phase (at 400 V)
Mains power connection
V/ Hz
192-254V/3PPE/60Hz
208-230/60/3 phases
208-230/60/3 phases
Current consumption without heater
A
20/ phase (at 230 V)
20/ phase (at 230 V)
Current consumption 5kW heater
A
21/ phase (at 230 V)
21/ phase (at 230 V)
All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.