Control parameters – xp, tv, tn – JULABO SC 10000w Eco-Series Recirculating Coolers User Manual
Page 37

Recirculating Cooler
37
Handshake
 
0 = Xon/Xoff protocol software handshake
1 = Protocol RTS/CTS
hardware handshake
 
 
Data bits = 7; Stop bits = 1
1.
Press the
button until the desired menu item is displayed.
2.
Press enter
to indicate the parameter.
3.
Select the parameter with the keys
.
4.
Press enter
to store the selected parameter.
9.5. Control parameters – Xp, Tv, Tn
The control parameters preset in factory are in most cases adequate for 
achieving an optimum temperature pattern for the samples requiring 
temperature application. 
Each parameter may be manually set via the keypad if necessary, to 
allow optimum control performance. 
 
Setting range: 
0.1 ... 99.9 
Proportional range >Xp<
The proportional range is the range below the selected temperature value 
in which the control circuit reduces the heating power from 100 % to 0 %.
 
Setting range:: 
1 ... 9999 
Resetting time >Tn< (Integral component)
Compensation of the remaining control deviation due to proportional 
regulation. An insufficient resetting time may cause instabilities to occur. 
Excessive resetting time will unnecessarily prolong compensation of the 
control difference. 
 
Setting range: 
0 ... 99 
Lead time >Tv< (Differential component)
The differential component reduces the control settling time. An 
insufficient lead time will prolong the time required to compensate for 
disturbance effects and cause high overshooting during run-up. An 
excessive lead time could cause instabilities (oscillations) to occur. 
 