Recirculating Cooler

35


Example

5.

Determine the calibration type:

(1-point, 2-point or 3-point calibration.)

Press the

button until the menu item

is displayed

and press enter

to indicate the parameter.

Select the parameter with the keys

(1, 2, 3).

Press enter

to store the selected parameter.

Depending on which calibration type is selected (1-point, 2-point or 3-
point calibration) only the necessary menu items are indicated.

For 1-point calibration:

>tt1< and >Ct1<

For 2-point calibration:

>tt1< and >Ct1<

>tt2<

and

>Ct2<

For 3-point calibration:

>tt1< and >Ct1<

>tt2<

and

>Ct2<

>tt3<

and

>Ct3<

From example on page 34
for T

M

= Ctx

= 79.7 °C

= 119.5 °C

= 159.3 °C


6.

Setting and storing the value:

Press the

button until the menu item

is displayed and press enter

to open the input

window.
Read the value of T

M

on the temperature measuring device and

enter the respective value.

Use the cursor keys to move left or right on the display
until the numeral you wish to change blinks.

Use the increase/decrease arrows

to change the

selected numeral (-, 0, 1, 2, 3, ... 9).

Press enter

and the circulator also stores the value of T

T

as value for

(80.00 °C).

The first of the 3 points is now calibrated.

The values for >ttx< can only be displayed, but they cannot be
modified.

7.

Repeat the calibration procedure for 120 °C and 160 °C.

Store the values under

or

resp.

Notice:

In the ATC status > YES<, the ATC calibration curve always affects the
effective working temperature (also the one set via the interface).