JULABO SC 10000w Eco-Series Recirculating Coolers User Manual

Page 34

background image

Menu functions

34

1.

2.

The menu ATC - Absolute Temperature Calibration
is structured in 2 levels.

ATC status
Calibration type: 1-, 2-, 3-point calibration

2 values per calibration point

ttx = Temperature indication on circulator
Indicates the value of the calibration point entered last.
Ctx = Calibration temperature x
Input of the new value for the calibration curve.



Examples:
T

T

= 80.0 °C

T

M

= 79.7 °C


T

T

= 120.0 °C

T

M

= 119.5 °C


T

T

= 160.0 °C

T

M

= 159.3 °C

Example:

3-point calibration

In the temperature range of 80 °C to 160 °C the calibration curve of the
temperature sensor (T

T

) should be assimilated to the actual

temperatures on the measuring point (T

M

).

1.

Calibration procedure:

Press

and set the first temperature value under e g.

for example (example 1

st

value = 80 °C).

2.

Wait until this temperature is maintained constant in the bath for
about 5 minutes.



ATC

status


Factory setting: YES




3.

Call up the >ATC< menu:

Press the

button until the menu item

is displayed

and press enter

to switch to menu level 2.

4.

Adjusting menu item >ATC Status<:

.

Press enter

and indicate the parameter.

Select the parameter with the keys

(no / YES).

Press enter

to store the selected parameter.

>no<

During the calibration process >no< needs to be set.

Continue: Press the

button to call up the menu

item

>Art<.


>YES<

The controller of the circulator uses the new calibration

curve. Set this parameter after the
calibration

process.

Continue:

Press

to quit the menu.