JULABO SC 10000w Eco-Series Recirculating Coolers User Manual
Page 31

Recirculating Cooler
31
8.5. Change-over of the warning function to shutdown function
 
Factory setting: Li 0 
 
 
For the two menu items > t High < and > t Low < choose between a 
warning message being signalled or a complete shutdown of the main 
functional elements such as heater and circulating pump being effected. 
(see page 29) 
 Setting >Li / 0<
An audible warning (interval tone) sounds and a meassage appears 
on the MULTI-DISPLAY (LED). 
t High 
t Low
or
 Setting >Li / 1<
A complete shutdown of heater and circulating pump is effected. 
An audible alarm (continuous tone) sounds and a message appears 
on the MULTI-DISPLAY (LED). 
t High 
t Low
or
.
1.
Press the key
until the menu item
is displayed.
2.
Press enter
to indicate the adjusted parameter on the MULTI-
DISPLAY (LED) (digit blinks).
3.
Select the parameter with the keys
(0 / 1).
4.
Press enter
to store the selected parameter.
Press
to update the display immediately.