Connecting the external system – JULABO SC 10000w Eco-Series Recirculating Coolers User Manual
Page 21
Recirculating Cooler
21
5.2. Connecting the external system
3/4" NPT male
0
1
24
19
Connect the tubing for cooling the external
system to the pump connectors (22, 23) for
feed and return on the rear of the recirculating
cooler.
In case the system to be cooled is located at
a higher level than the recirculating cooler,
take note of bath liquid flowing back when the
unit is switched off. Should the filling volume
of the bath tank not be sufficient, prevent the
liquid from flowing back by using shut-off
valves.
Flood hazard!
For this reason, solenoid valves for loop
circuit or shut-off valves can be integrated in
the loop circuit.
Connect the valve to the connector (19).
Order No.
Description
8 980 705 Solenoid valve set 230V/50-60Hz
Unscrew the collar nut from the overflow connector (24).
Connect a piece of tubing to the overflow connector and drain into
a suitable vessel.
Order-No. 8 970 460
Barbed fitting for tubing 8 mm inner dia.
If easily volatile temperature liquids are used, as e.g. 3M
Fluorinert
®
, the overflow should remain closed.
Attention Flood hazard!
Then the level indication (6) should get more attention.
Caution:
Securely attach all tubing to prevent slipping.